Detectors:
Scintillator: CsI(Na), 9 mm thick 3?5” curved array, 2.2?2.2 mm crystals
Intrinsic energy resolution: 30%

Tunable slit-rake collimators:
Slit opening range: 0 to 5 mm
Slit material: Iridium
Longitudinal rake collimation: 0.3 mm thick tungsten sheets, 2.2 mm equally spaced
Focal length: 65 mm

Measured transverse spatial resolution: 0.35 mm hot rods and 0.65 mm cold rods (32 mm outer diameter phantom)
Longitudinal spatial resolution: 0.60 mm (interleave acquisition)
Measured sensitivity: 245 cpm/?Ci (0.2 mm slit opening, 32 mm outer diameter phantom)
Maximal animal diameter: 9 cm
Longitudinal field of view: 7.5 cm, Bed length: 25 cm
Anti collimator collision system

Computer:
Acquisition: pentium computer. Processing: 4-xeon parallelized computer.

Software:
Full list mode acquisition: recording of the (x,y) coordinate, time and energy per event
Rebining: static, dynamic, multi-bed, gated and interleave tomographic acquisition
Processing: FBP and EM-ML reconstruction, orthogonal viewer, interfile and Dicom export files

Linoview™miniSPECT will be supplied as a mobile device adapted to your preferences.